Global Medical Directory is part of the Informa Markets Division of Informa PLC

This site is operated by a business or businesses owned by Informa PLC and all copyright resides with them. Informa PLC's registered office is 5 Howick Place, London SW1P 1WG. Registered in England and Wales. Number 8860726.

B-2000 Infant Incubator
Connect with this vendor - Learn about this product

B-2000 Infant Incubator

STANDARD FEATURES:

●Servo Control Mode: air temperature, skintemperature;

● Parameter Displaying: set temperature,air temperature, skin temperature, internal humidity, heat power rate;

● Audible & Visual Alarms: powerfailure, sensor failure, higher & lower temp, air & skin temperature override,fan failure, system failure;

● >37℃Preset Mode: power failure reset memory, alarm silence, keypad lock, self-testfunction;

● Hood with dual slope; 6 hand-ports;

● RS-232 Output Connector (use for networksystem);

●IV poles;

●Protective guard rail;

●Storage cabinets.

 

 

Specifications of hood

Mattress to Hood: 450 mm

Observation side of hood: Two incline

Operate port: 6 (2 Iris ports)

Tubing grommets: 4

Front access panel: 826×310 mm

Mattress Tray Size: 630×360 mm

Technical Specifications:    

Powerrequirement ≤400VA

Temperaturecontrol mode: Air and skin servo control

AirTemp Control Range: 20℃~37℃

Temperatureoverride mode to: 20℃~39℃

SkinTemp Control Range: 30℃~37℃

Temperatureoverride mode to: 30℃~38℃

Skintemp sensor accuracy: ≤0.3℃

Temperaturefluctuation: ≤0.5℃

Mattresstemperature uniformity: ≤0.8℃

Temperaturerise time: ≤45 min

Noiselevel within hood environment: ≤55dB

Alarms  

Power failure alarm        

Fan failure alarm

Sensor failure alarm

High air temperature alarm: +3.0℃

Low airtemperature alarm: -3.0℃

High skin temperature alarm: +1.0℃

Low skin temperature alarm: -1.0℃

Over air temperature alarm: ≤38.0℃; ≤40.0℃

Over skin temperature alarm: ≤38.0℃; ≤39.0℃

Systemfailure alarm  

Alarmsilence function

Operational condition  

Environment temperature: +20℃~+30℃

Environment relative humidity: 5%~99%

Environment air velocity of flow: <0.3m/s

Others    

Infant bed tilt adjusting  ±10° step-less

Mattress tray Max. load  10kg

Transfusion pole Max. load   2kg

CabinetMax. load  6kg

Specialty: