Global Medical Directory is part of the Informa Markets Division of Informa PLC

This site is operated by a business or businesses owned by Informa PLC and all copyright resides with them. Informa PLC's registered office is 5 Howick Place, London SW1P 1WG. Registered in England and Wales. Number 8860726.

B-800 Infant Incubator
Connect with this vendor - Learn about this product

B-800 Infant Incubator

STANDARD COMPOSITION:

● Servo Control Mode: airtemp;

● Instant ParameterDisplaying: set temp, air temp, heat power rate;

● Audible & Visual Alarms:power failure, sensor failure, higher/lower temp, air temperature override, fan failure;

● Power failure reset memory and self-test function;

● Hood with dual slope: 4hand-ports.

Specifications of hood

Mattress to Hood: 450 mm

Observation side of hood: Two incline

Operate port: 4 (2 Iris ports)

Tubing grommets: 2

Front access panel: 826×310 mm

Mattress Tray Size: 630×360 mm

 

Technical Specifications:    

Power requirement ≤400VA

Temperature control mode: Air servo control

Air Temp Control Range: 20℃~37℃

Temperature fluctuation: ≤0.5℃

Mattress temperature uniformity: ≤0.8℃

Temperature rise time: ≤45 min

Noise level within hood environment: ≤55dB

 

Alarms  

Power failure alarm        

Fan failure alarm

Sensor failure alarm

High air temperature alarm: +3.0℃

Low air temperature alarm: -3.0℃

Over air temperature alarm: ≤38.0℃

Operational condition  

Environment temperature: +20℃~+30℃

Environment relative humidity: 5%~99%

Environment air velocity of flow: <0.3m/s

 

Others    

Mattress tray Max. load  10kg

Transfusion pole Max. load   2kg

Specialty: