MedicCleanAir - Isolation Unit - ISO 210
This unit is made for installation against the wall (ISO210), does not take any floor space. Easy and quick installation (1/2 day time) and functions independently of the existing ventilation (A/C) system.
The unit consists of two pieces; filter part + electronics and ventilator. Both parts include pressure sensors which lead to permanent pressure difference monitoring (optional) between the patient’s room and the surrounding area e.g. the corridor. The device is provided with an air back-pressure flap that avoids return of airflow.
When the room is completely sealed off, one may expect pressure differences up to 20 Pascal. Max capacity 135m3/hour, for rooms of 300cm*200cm*H250cm
To create negative or positive pressure in larger rooms we recommend the MedicCleanAir ISO520 device.